Purpose: To develop and apply a stepping approach for the validation of Knowledge-based (KB) models for planning optimization: the method was applied to the case of concomitant irradiation of pelvic nodes and prostate + seminal−vesicles bed irradiation in post-prostatectomy patients. Methods: The clinical VMAT plans of 52 patients optimized by two reference planners were selected to generate a KB-model (RapidPlan, v.13.5 Varian). A stepping-validation approach was followed by comparing KB-generated plans (with and without planner-interaction, RP and only-RP respectively) against delivered clinical plans (RA). The validation followed three steps, gradually extending its generalization: 20 patients used to develop the model (closed-loop); 20 new patients, same planners (open-loop); 20 new patients, different planners (wide-loop). All plans were compared, in terms of relevant dose-volume parameters and generalized equivalent uniform dose (gEUD). Results: KB-plans were generally better than or equivalent to clinical plans. For RPvsRA, PTVs coverage was comparable, for OARs RP was always better. Comparing only-RPvsRA, PTVs coverage was always better; bowelladder V 50Gy and D 1% , bowelladderectum D mean , femoral heads V 40Gy and penile bulb V 50Gy were significantly improved. For RPvsRA gEUD reduction >1 Gy was seen in 80% of plans for rectum, bladder and bowel; for only-RPvsRA, this was found in 50% for rectum/bladder and in 70% for bowel. Conclusion: An extensive stepping validation approach of KB-model for planning optimization showed better or equal performances of automatically generated KB-plan compared to clinical plans. The interaction of a planner further improved planning performances.

Comprehensive Intra-Institution stepping validation of knowledge-based models for automatic plan optimization / Castriconi, R.; Fiorino, C.; Broggi, S.; Cozzarini, C.; Di Muzio, N.; Calandrino, R.; Cattaneo, G. M.. - In: PHYSICA MEDICA. - ISSN 1120-1797. - 57:(2019), pp. 231-237. [10.1016/j.ejmp.2018.12.002]

Comprehensive Intra-Institution stepping validation of knowledge-based models for automatic plan optimization

Di Muzio N.;
2019-01-01

Abstract

Purpose: To develop and apply a stepping approach for the validation of Knowledge-based (KB) models for planning optimization: the method was applied to the case of concomitant irradiation of pelvic nodes and prostate + seminal−vesicles bed irradiation in post-prostatectomy patients. Methods: The clinical VMAT plans of 52 patients optimized by two reference planners were selected to generate a KB-model (RapidPlan, v.13.5 Varian). A stepping-validation approach was followed by comparing KB-generated plans (with and without planner-interaction, RP and only-RP respectively) against delivered clinical plans (RA). The validation followed three steps, gradually extending its generalization: 20 patients used to develop the model (closed-loop); 20 new patients, same planners (open-loop); 20 new patients, different planners (wide-loop). All plans were compared, in terms of relevant dose-volume parameters and generalized equivalent uniform dose (gEUD). Results: KB-plans were generally better than or equivalent to clinical plans. For RPvsRA, PTVs coverage was comparable, for OARs RP was always better. Comparing only-RPvsRA, PTVs coverage was always better; bowelladder V 50Gy and D 1% , bowelladderectum D mean , femoral heads V 40Gy and penile bulb V 50Gy were significantly improved. For RPvsRA gEUD reduction >1 Gy was seen in 80% of plans for rectum, bladder and bowel; for only-RPvsRA, this was found in 50% for rectum/bladder and in 70% for bowel. Conclusion: An extensive stepping validation approach of KB-model for planning optimization showed better or equal performances of automatically generated KB-plan compared to clinical plans. The interaction of a planner further improved planning performances.
2019
Automatic planning
Knowledge-based
Machine learning
Prostate radiotherapy
Automation
Radiotherapy Planning, Computer-Assisted
Models, Theoretical
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.11768/108090
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 19
  • ???jsp.display-item.citation.isi??? 20
social impact