We evaluated the safety and effectiveness of postdilating a 3.0 mm sirolimus-eluting stents (SESs; six cells) with a 3.5-4.0 mm balloon. We identified 254 consecutive patients who underwent percutaneous coronary interventions using SESs with a nominal diameter of 3.0 mm (six cells). Patients were divided into two groups based on whether they were subsequently postdilated with a 3.0 mm (group 1: 168 patients, 251 lesions) or a 3.5-4 mm balloon (group 2: 86 patients, 102 lesions). There were no significant differences regarding the incidence of in-hospital and long-term follow-up. Angiographic follow-up was available in 72% and 74% of groups 1 and 2, respectively. The two groups had no significant differences regarding late lumen loss (0.51 +/- 0.36 vs. 0.52 +/- 0.33; P = 0.3) and binary restenosis rates (10.7% vs. 8.8%; P = 0.1). Six-month clinical follow-up was available in all patients. At long-term follow-up (mean: 10.6 +/- 3.7 for group 1 and 11.3 +/- 3.9 months for group 2), there were no significant differences between the two groups regarding major adverse cardiac events (8.9% vs. 9.2%; P = 0.9). Implantation of a 3.0 mm SES with postdilation with a 3.5-4 mm balloon did not result in any significant difference in complications, in-hospital non-Q-wave myocardial infarction, binary restenosis, or target lesion revascularization. These data should lessen concern that overdilation may dilute the beneficial effects of SESs.

Is overdilatation of 3.0 mm sirolimus-eluting stent associated with a higher restenosis rate? / Iakovou, Ioannis; Stankovic, Goran; Montorfano, M; Airoldi, Flavio; Chieffo, Alaide; Sangiorgi Giuseppe, M.; Carlino, Mauro; Corvaja, Nicola; Iassen, Michev; Rogacka, Renata; Vitrella, Giancarlo; Colombo, Antonio. - In: CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS. - ISSN 1522-1946. - (2005). [10.1002/ccd.20276]

Is overdilatation of 3.0 mm sirolimus-eluting stent associated with a higher restenosis rate?

Montorfano M;Chieffo Alaide;
2005-01-01

Abstract

We evaluated the safety and effectiveness of postdilating a 3.0 mm sirolimus-eluting stents (SESs; six cells) with a 3.5-4.0 mm balloon. We identified 254 consecutive patients who underwent percutaneous coronary interventions using SESs with a nominal diameter of 3.0 mm (six cells). Patients were divided into two groups based on whether they were subsequently postdilated with a 3.0 mm (group 1: 168 patients, 251 lesions) or a 3.5-4 mm balloon (group 2: 86 patients, 102 lesions). There were no significant differences regarding the incidence of in-hospital and long-term follow-up. Angiographic follow-up was available in 72% and 74% of groups 1 and 2, respectively. The two groups had no significant differences regarding late lumen loss (0.51 +/- 0.36 vs. 0.52 +/- 0.33; P = 0.3) and binary restenosis rates (10.7% vs. 8.8%; P = 0.1). Six-month clinical follow-up was available in all patients. At long-term follow-up (mean: 10.6 +/- 3.7 for group 1 and 11.3 +/- 3.9 months for group 2), there were no significant differences between the two groups regarding major adverse cardiac events (8.9% vs. 9.2%; P = 0.9). Implantation of a 3.0 mm SES with postdilation with a 3.5-4 mm balloon did not result in any significant difference in complications, in-hospital non-Q-wave myocardial infarction, binary restenosis, or target lesion revascularization. These data should lessen concern that overdilation may dilute the beneficial effects of SESs.
2005
Pubblicato
Is overdilatation of 3.0 mm sirolimus-eluting stent associated with a higher restenosis rate? / Iakovou, Ioannis; Stankovic, Goran; Montorfano, M; Airoldi, Flavio; Chieffo, Alaide; Sangiorgi Giuseppe, M.; Carlino, Mauro; Corvaja, Nicola; Iassen, Michev; Rogacka, Renata; Vitrella, Giancarlo; Colombo, Antonio. - In: CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS. - ISSN 1522-1946. - (2005). [10.1002/ccd.20276]
none
12
info:eu-repo/semantics/article
262
Iakovou, Ioannis; Stankovic, Goran; Montorfano, M; Airoldi, Flavio; Chieffo, Alaide; Sangiorgi Giuseppe, M.; Carlino, Mauro; Corvaja, Nicola; Iassen, ...espandi
1 Contributo su Rivista::1.1 Articolo in rivista
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.11768/142987
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 18
  • ???jsp.display-item.citation.isi??? 11
social impact