Introduction: This study compared the canal curvature and axis modification after instrumentation with WaveOne Primary reciprocating files (Dentsply Maillefer, Ballaigues, Switzerland) and nickel-titanium (NiTi) rotary ProTaper (Dentsply Maillefer). Methods: Thirty ISO 15, 0.02 taper, Endo Training Blocks (Dentsply Maillefer) were used. In all specimens, the glide path was achieved with PathFile 1, 2, and 3 (Dentsply Maillefer) at the working length (WL). Specimens were then assigned to 1 of 2 groups for shaping: specimens in group 1 were shaped with ProTaper S1-S2-F1-F2 at the WL and specimens in group 2 were shaped with WaveOne Primary reciprocating files at the WL. Pre- and postinstrumentation digital images were superimposed and processed with Matlab r2010b (The MathWorks Inc, Natick, MA) software to analyze the curvature-radius ratio (CRr) and the relative axis error (rAe), representing canal curvature modification. Data were analyzed with one-way balanced analyses of variance at 2 levels (P <.05). Results: The instrument factor was extremely significant for both the CRr parameter (F1 = 9.59, P =.004) and the rAe parameter (F1 = 13.55, P =.001). Conclusions: Canal modifications are reduced when the new WaveOne NiTi single-file system is used. © 2012 American Association of Endodontists.

Canal shaping with waveone primary reciprocating files and protaper system: A comparative study / Berutti, E.; Chiandussi, G.; Paolino, D. S.; Scotti, N.; Cantatore, G.; Castellucci, A.; Pasqualini, D.. - In: JOURNAL OF ENDODONTICS. - ISSN 0099-2399. - 38:4(2012), pp. 505-509. [10.1016/j.joen.2011.12.040]

Canal shaping with waveone primary reciprocating files and protaper system: A comparative study

Cantatore G.
Investigation
;
2012-01-01

Abstract

Introduction: This study compared the canal curvature and axis modification after instrumentation with WaveOne Primary reciprocating files (Dentsply Maillefer, Ballaigues, Switzerland) and nickel-titanium (NiTi) rotary ProTaper (Dentsply Maillefer). Methods: Thirty ISO 15, 0.02 taper, Endo Training Blocks (Dentsply Maillefer) were used. In all specimens, the glide path was achieved with PathFile 1, 2, and 3 (Dentsply Maillefer) at the working length (WL). Specimens were then assigned to 1 of 2 groups for shaping: specimens in group 1 were shaped with ProTaper S1-S2-F1-F2 at the WL and specimens in group 2 were shaped with WaveOne Primary reciprocating files at the WL. Pre- and postinstrumentation digital images were superimposed and processed with Matlab r2010b (The MathWorks Inc, Natick, MA) software to analyze the curvature-radius ratio (CRr) and the relative axis error (rAe), representing canal curvature modification. Data were analyzed with one-way balanced analyses of variance at 2 levels (P <.05). Results: The instrument factor was extremely significant for both the CRr parameter (F1 = 9.59, P =.004) and the rAe parameter (F1 = 13.55, P =.001). Conclusions: Canal modifications are reduced when the new WaveOne NiTi single-file system is used. © 2012 American Association of Endodontists.
2012
Canal shaping
nickel-titanium
ProTaper
reciprocating motion
WaveOne
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.11768/159056
Citazioni
  • ???jsp.display-item.citation.pmc??? 50
  • Scopus 149
  • ???jsp.display-item.citation.isi??? 103
social impact