The aim of this study was to compare cardiac volume and function assessment using PET with the reference technique of cardiovascular magnetic resonance (CMR). Methods: Left ventricular (LV) and right ventricular (RV) end-diastolic volume (EDV), end-systolic volume (ESV), stroke volume (SV), and ejection fractions (EF) were measured in 9 patients using both CMR and PET with inhaled (CO)-O-15. Results: Correlation between the techniques was generally reasonable (r values ranged from 0.63 to 0.99). Best agreement was seen for ESV (LV and RV). With PET, there was a tendency to underestimate LV EF and EDV, and RV EDV and SV. Agreement was worst for LV SV. Percentage difference between CMR and PET measurements ranged from -2% to 15%; Bland-Altman limits of agreement ranged from 24% to 75%. Conclusion: Although small systematic differences exist, the agreement between PET and CMR suggests useful information regarding function, and volumes may be obtained from a standard PET protocol.

Measurement of ventricular volumes and function: A comparison of gated PET and cardiovascular magnetic resonance

CAMICI , PAOLO;
2002-01-01

Abstract

The aim of this study was to compare cardiac volume and function assessment using PET with the reference technique of cardiovascular magnetic resonance (CMR). Methods: Left ventricular (LV) and right ventricular (RV) end-diastolic volume (EDV), end-systolic volume (ESV), stroke volume (SV), and ejection fractions (EF) were measured in 9 patients using both CMR and PET with inhaled (CO)-O-15. Results: Correlation between the techniques was generally reasonable (r values ranged from 0.63 to 0.99). Best agreement was seen for ESV (LV and RV). With PET, there was a tendency to underestimate LV EF and EDV, and RV EDV and SV. Agreement was worst for LV SV. Percentage difference between CMR and PET measurements ranged from -2% to 15%; Bland-Altman limits of agreement ranged from 24% to 75%. Conclusion: Although small systematic differences exist, the agreement between PET and CMR suggests useful information regarding function, and volumes may be obtained from a standard PET protocol.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.11768/1658
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 45
  • ???jsp.display-item.citation.isi??? 40
social impact