A total of 1001 subjects (82.0 years, 62.5% female, 63.7% NYHA III/IV at baseline) with severe aortic stenosis at high surgical risk were enrolled in the prospective CONFIDENCE registry and treated with a Portico™ transcatheter heart valve (THV) using either a first-generation delivery system (DS) or the FlexNav™ DS. The objective of this registry is to characterize the procedural safety and device performance of the Portico™ THV at 30 days. The study collected ‘standard-of-care’ clinical and device performance data, with adverse events adjudicated by an independent clinical event committee according to the Valve Academic Research Consortium-2 criteria. The implantation of a single Portico™ THV was successful in 97.5% of subjects. The 30-day all-cause mortality, cardiovascular mortality, and disabling stroke rates were 2.6%, 2.1%, and 1.8%, respectively. A new pacemaker was implanted in 19.0% of subjects at 30 days. At 30 days, the effective orifice area and mean gradient values were 1.82 cm2 and 7.1 mmHg, respectively. The 30-day rate of moderate paravalvular leak (PVL) was 2.1%, with no occurrence of severe PVL. The Portico™ THV demonstrated improved hemodynamic performance and low rates of safety events at 30 days in a large cohort of subjects implanted with the Portico™ THV with either the first-generation DS or FlexNav™ DS.

Procedural Safety and Device Performance of the Portico™ Valve from Experienced TAVI Centers: 30-Day Outcomes in the Multicenter CONFIDENCE Registry / Mollmann, H.; Linke, A.; Nombela-Franco, L.; Sluka, M.; Dominguez, J. F. O.; Montorfano, M.; Kim, W. -K.; Arnold, M.; Vasa-Nicotera, M.; Conradi, L.; Camuglia, A.; Bedogni, F.; Manoharan, G.. - In: JOURNAL OF CLINICAL MEDICINE. - ISSN 2077-0383. - 11:16(2022). [10.3390/jcm11164839]

Procedural Safety and Device Performance of the Portico™ Valve from Experienced TAVI Centers: 30-Day Outcomes in the Multicenter CONFIDENCE Registry

Montorfano M.;
2022-01-01

Abstract

A total of 1001 subjects (82.0 years, 62.5% female, 63.7% NYHA III/IV at baseline) with severe aortic stenosis at high surgical risk were enrolled in the prospective CONFIDENCE registry and treated with a Portico™ transcatheter heart valve (THV) using either a first-generation delivery system (DS) or the FlexNav™ DS. The objective of this registry is to characterize the procedural safety and device performance of the Portico™ THV at 30 days. The study collected ‘standard-of-care’ clinical and device performance data, with adverse events adjudicated by an independent clinical event committee according to the Valve Academic Research Consortium-2 criteria. The implantation of a single Portico™ THV was successful in 97.5% of subjects. The 30-day all-cause mortality, cardiovascular mortality, and disabling stroke rates were 2.6%, 2.1%, and 1.8%, respectively. A new pacemaker was implanted in 19.0% of subjects at 30 days. At 30 days, the effective orifice area and mean gradient values were 1.82 cm2 and 7.1 mmHg, respectively. The 30-day rate of moderate paravalvular leak (PVL) was 2.1%, with no occurrence of severe PVL. The Portico™ THV demonstrated improved hemodynamic performance and low rates of safety events at 30 days in a large cohort of subjects implanted with the Portico™ THV with either the first-generation DS or FlexNav™ DS.
2022
aortic stenosis
Portico
self-expanding
transcatheter aortic valve implantation
transcatheter aortic valve replacement
File in questo prodotto:
File Dimensione Formato  
jcm-11-04839.pdf

accesso aperto

Tipologia: PDF editoriale (versione pubblicata dall'editore)
Licenza: Creative commons
Dimensione 866.62 kB
Formato Adobe PDF
866.62 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.11768/170478
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 13
  • ???jsp.display-item.citation.isi??? 12
social impact