A total of 1001 subjects (82.0 years, 62.5% female, 63.7% NYHA III/IV at baseline) with severe aortic stenosis at high surgical risk were enrolled in the prospective CONFIDENCE registry and treated with a Portico™ transcatheter heart valve (THV) using either a first-generation delivery system (DS) or the FlexNav™ DS. The objective of this registry is to characterize the procedural safety and device performance of the Portico™ THV at 30 days. The study collected ‘standard-of-care’ clinical and device performance data, with adverse events adjudicated by an independent clinical event committee according to the Valve Academic Research Consortium-2 criteria. The implantation of a single Portico™ THV was successful in 97.5% of subjects. The 30-day all-cause mortality, cardiovascular mortality, and disabling stroke rates were 2.6%, 2.1%, and 1.8%, respectively. A new pacemaker was implanted in 19.0% of subjects at 30 days. At 30 days, the effective orifice area and mean gradient values were 1.82 cm2 and 7.1 mmHg, respectively. The 30-day rate of moderate paravalvular leak (PVL) was 2.1%, with no occurrence of severe PVL. The Portico™ THV demonstrated improved hemodynamic performance and low rates of safety events at 30 days in a large cohort of subjects implanted with the Portico™ THV with either the first-generation DS or FlexNav™ DS.
Procedural Safety and Device Performance of the Portico™ Valve from Experienced TAVI Centers: 30-Day Outcomes in the Multicenter CONFIDENCE Registry / Mollmann, H.; Linke, A.; Nombela-Franco, L.; Sluka, M.; Dominguez, J. F. O.; Montorfano, M.; Kim, W. -K.; Arnold, M.; Vasa-Nicotera, M.; Conradi, L.; Camuglia, A.; Bedogni, F.; Manoharan, G.. - In: JOURNAL OF CLINICAL MEDICINE. - ISSN 2077-0383. - 11:16(2022). [10.3390/jcm11164839]
Procedural Safety and Device Performance of the Portico™ Valve from Experienced TAVI Centers: 30-Day Outcomes in the Multicenter CONFIDENCE Registry
Montorfano M.;
2022-01-01
Abstract
A total of 1001 subjects (82.0 years, 62.5% female, 63.7% NYHA III/IV at baseline) with severe aortic stenosis at high surgical risk were enrolled in the prospective CONFIDENCE registry and treated with a Portico™ transcatheter heart valve (THV) using either a first-generation delivery system (DS) or the FlexNav™ DS. The objective of this registry is to characterize the procedural safety and device performance of the Portico™ THV at 30 days. The study collected ‘standard-of-care’ clinical and device performance data, with adverse events adjudicated by an independent clinical event committee according to the Valve Academic Research Consortium-2 criteria. The implantation of a single Portico™ THV was successful in 97.5% of subjects. The 30-day all-cause mortality, cardiovascular mortality, and disabling stroke rates were 2.6%, 2.1%, and 1.8%, respectively. A new pacemaker was implanted in 19.0% of subjects at 30 days. At 30 days, the effective orifice area and mean gradient values were 1.82 cm2 and 7.1 mmHg, respectively. The 30-day rate of moderate paravalvular leak (PVL) was 2.1%, with no occurrence of severe PVL. The Portico™ THV demonstrated improved hemodynamic performance and low rates of safety events at 30 days in a large cohort of subjects implanted with the Portico™ THV with either the first-generation DS or FlexNav™ DS.File | Dimensione | Formato | |
---|---|---|---|
jcm-11-04839.pdf
accesso aperto
Tipologia:
PDF editoriale (versione pubblicata dall'editore)
Licenza:
Creative commons
Dimensione
866.62 kB
Formato
Adobe PDF
|
866.62 kB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.