This article reports an evaluation of the Coulter Counter model S-Plus VI automatic analyser for haematology, and data are presented on linearity, carry-over, precision, accuracy and stability of the instrument, when compared with a model S-Plus IV/D.The three-part differential count provided by Coulter S-Plus VI has been compared with manual eye counting. The results show a good agreement with only 2.5% of discrepancies in 2271 routine samples.ADVANTAGES OF THE NEW INSTRUMENT INCLUDE: reduction of running costs, largely due to manpower saving; simple and easy use, and improved operator safety, there being no need for human contact with blood.

Evaluation of the Coulter Counter S-Plus VI

BANFI , GIUSEPPE;
1988-01-01

Abstract

This article reports an evaluation of the Coulter Counter model S-Plus VI automatic analyser for haematology, and data are presented on linearity, carry-over, precision, accuracy and stability of the instrument, when compared with a model S-Plus IV/D.The three-part differential count provided by Coulter S-Plus VI has been compared with manual eye counting. The results show a good agreement with only 2.5% of discrepancies in 2271 routine samples.ADVANTAGES OF THE NEW INSTRUMENT INCLUDE: reduction of running costs, largely due to manpower saving; simple and easy use, and improved operator safety, there being no need for human contact with blood.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.11768/1722
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 3
  • ???jsp.display-item.citation.isi??? 2
social impact