The impact of hemolysis during microaxial flow pump (mAFP; Impella, Danvers, Massachusetts, US) support on early outcomes after durable left ventricular assist device (d-LVAD) implantation is unknown. Three hundred and eleven consecutive patients undergoing d-LVAD implantation after mAFP support (Impella 5.0/5.5 72.3%) were retrospectively included. The incidence and predictors of hemolysis (plasma-free hemoglobin >20 mg/dl or lactic dehydrogenase (LDH) >2.5-fold the upper reference limit) before d-LVAD implantation were assessed, along with its impact on early post-d-LVAD outcomes. The primary outcome was a composite of hemocompatibility-related adverse events (HRAEs: stroke/gastrointestinal bleeding/ pump thrombosis). Hemolysis occurred in 40.8%. Impella 2.5/CP versus 5.0/5.5 was the single independent predictor of hemolysis (adj-hazard ratio [HR] = 2.68, 95% confidence interval [CI] = 1.04-6.94, p = 0.031). Post-d-LVAD HRAEs occurred more frequently among patients with hemolysis (31.9% vs. 20.6%; p = 0.041), mainly driven by hemorrhagic stroke and gastrointestinal bleeding. At multivariate analysis, hemolysis remained independently associated with HRAEs (adj-HR = 1.62, 95% CI = 1.02-2.58; p = 0.041). Patients with hemolysis were more likely to need a temporary right ventricular assist device following d-LVAD implantation (28.3% vs. 16.8%; p = 0.012), with no difference in mortality (23.6% vs. 21.2%; p = 0.355). In conclusion, among patients undergoing d-LVAD implantation with mAFP bridge, hemolysis is common, occurs more frequently among patients supported with Impella 2.5/CP, and is an independent predictor of post-d-LVAD HRAEs.

Impact of Hemolysis During Microaxial Flow Pump Support on Early Outcomes After Durable Left Ventricular Assist Device Implantation / Gallone, G., Lewin, D., Spitaleri, A., Hernandez, S.R., Bernhardt, A., Billion, M., Meyer, A., Netuka, I., Kooij, J., Pieri, M., Szymanski, M., Moeller, C., Akhyari, P., Jawad, K., Krasivskyi, I., Schmack, B., Farber, G., Medina, M., Haneya, A., Zimpfer, D., et al.. - In: ASAIO JOURNAL. - ISSN 1058-2916. - (2025). [10.1097/MAT.0000000000002451]

Impact of Hemolysis During Microaxial Flow Pump Support on Early Outcomes After Durable Left Ventricular Assist Device Implantation

Pieri M.;
2025-01-01

Abstract

The impact of hemolysis during microaxial flow pump (mAFP; Impella, Danvers, Massachusetts, US) support on early outcomes after durable left ventricular assist device (d-LVAD) implantation is unknown. Three hundred and eleven consecutive patients undergoing d-LVAD implantation after mAFP support (Impella 5.0/5.5 72.3%) were retrospectively included. The incidence and predictors of hemolysis (plasma-free hemoglobin >20 mg/dl or lactic dehydrogenase (LDH) >2.5-fold the upper reference limit) before d-LVAD implantation were assessed, along with its impact on early post-d-LVAD outcomes. The primary outcome was a composite of hemocompatibility-related adverse events (HRAEs: stroke/gastrointestinal bleeding/ pump thrombosis). Hemolysis occurred in 40.8%. Impella 2.5/CP versus 5.0/5.5 was the single independent predictor of hemolysis (adj-hazard ratio [HR] = 2.68, 95% confidence interval [CI] = 1.04-6.94, p = 0.031). Post-d-LVAD HRAEs occurred more frequently among patients with hemolysis (31.9% vs. 20.6%; p = 0.041), mainly driven by hemorrhagic stroke and gastrointestinal bleeding. At multivariate analysis, hemolysis remained independently associated with HRAEs (adj-HR = 1.62, 95% CI = 1.02-2.58; p = 0.041). Patients with hemolysis were more likely to need a temporary right ventricular assist device following d-LVAD implantation (28.3% vs. 16.8%; p = 0.012), with no difference in mortality (23.6% vs. 21.2%; p = 0.355). In conclusion, among patients undergoing d-LVAD implantation with mAFP bridge, hemolysis is common, occurs more frequently among patients supported with Impella 2.5/CP, and is an independent predictor of post-d-LVAD HRAEs.
2025
bridge strategy
hemolysis
impella
left ventricular assist device
microaxial flow pump
outcomes
File in questo prodotto:
File Dimensione Formato  
impact_of_hemolysis_during_microaxial_flow_pump.697.pdf

solo gestori archivio

Tipologia: PDF editoriale (versione pubblicata dall'editore)
Licenza: Tutti i diritti riservati
Dimensione 809.36 kB
Formato Adobe PDF
809.36 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.11768/189838
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 6
  • ???jsp.display-item.citation.isi??? 6
social impact