No clinical standardized methods exist to identify the early stage of the development of pump thrombosis in the setting of HVAD (Medtronic Inc., USA) implantation. We aimed at developing a clinically relevant tool to evaluate HVAD operation during long-term support and at identifying a new reliable marker for the early diagnosis of pump thrombosis reflecting altered patient-pump physiological interplay.
Log files analysis and evaluation of circadian patterns for the early diagnosis of pump thrombosis with a centrifugal continuous-flow left ventricular assist device / Consolo, Filippo; Esposti, Federico; Gustar, Adrian; De Bonis, Michele; Pappalardo, Federico. - In: THE JOURNAL OF HEART AND LUNG TRANSPLANTATION. - ISSN 1053-2498. - (2019). [Epub ahead of print] [10.1016/j.healun.2019.04.008]
Log files analysis and evaluation of circadian patterns for the early diagnosis of pump thrombosis with a centrifugal continuous-flow left ventricular assist device
Consolo, Filippo
Primo
;Esposti, Federico;De Bonis, Michele;Pappalardo, Federico
2019-01-01
Abstract
No clinical standardized methods exist to identify the early stage of the development of pump thrombosis in the setting of HVAD (Medtronic Inc., USA) implantation. We aimed at developing a clinically relevant tool to evaluate HVAD operation during long-term support and at identifying a new reliable marker for the early diagnosis of pump thrombosis reflecting altered patient-pump physiological interplay.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.